COMPARISON OF ENERGY DOSES 10 MV DISTRIBUTION USING PERCENTAGE DEPTH DOSE (PDD) METHOD ON LINAC: ELECTA AND SIEMENS

Martha Rianna, Herty Afrina Sianturi, Hariyati Lubis, Awan Pelawi, Timbangen Sembiring, Marhaposan Situmorang

Abstract


The patient dosing on Linac (Electa and Siemens) can be determined by Source Surface Distance (SSD) technique using Precentage Depth Doses (PDD) method. The study was conducted by measuring PDD to compare the dosage distribution calculations on Linac Electa and Siemens device of photon energy at 10 MV. PDD is done with a 100 cm SSD technique at a depth of 0 to 25 cm. The dose distribution results between the Electrical and the Siemens PDD are almost the same in that the Dmax at 10 MV Siemens photon energy occurs at a depth of 20 mm while the 10 MV Electa photon energy occurs at a depth of 21 mm. Both Linac Electa and Siemens device this at the same energy of 10 MV there is a difference of 95.23%.

Keywords: Distrubusi dose, PDD, Photon Energy, Quality File Index


Keywords


Distrubusi dose, PDD, Photon Energy, Quality File Index

Full Text:

PDF

References


Podgorsak E.B., 2005, Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency, Technical Editor, Vienna. Thirth Edition.(179-191)

AAPM (American Association of Physicists in Medicine) Report No. 13, 1994. “Physical Ascpects of Quality Assurance in Radioation Therapy”, AAPM, New York.

Sani, G., 2001. “Radiation Dosimetry Instrument and Methods, Second Edition”,CRC Press.

Chen C., 2007, Principles and requirements of external beam dosimetry, Elselvier, Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut, USA. S2 – S21

Podgorsak, E. B., 2005, Eksternal Photon Beams : Physical Aspect in Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency, Vienna : Publishing Section IAEA.

Kriˇzan P., Rihard H., 2008, Teleradiography Dose Calculation, Faculty of Mathematics and Physics, University of Ljubljana.

Khan F. M., 2003, Title : Physics of Radiation Therapy, The 3rd Edition, Lippincott Williams & Wilkins, Philadelphia, USA(136-137,164)

Buzdar S. A., Rao A., Aalia N., 2009, An Analyis 0f Depth Dose Charateristik of Photon in Water, Journal Ayub Med Coll Abbottabad, The Islamia University, Bahawalpur, Pakistan. 21(4)

Alam M. J., Rabbani Z., Hussain A., Baig A. K. V., 2007, A modified formula for defining tissue phantom ratio of photon beams, Bangladesh Medical Research Council. 33: 92-97

Richmond N., Robert B., 2014, A Comparison of Small-Field Tissue Phantom Ratio Data Generation Methods for an Electa Agility 6 MV Photon Beam, Elsevier, Medical Physics Department, The James Cook University Hospital, Middlesbrough, United Kingdom.60-63

Sianturi, Herty Afrina., Martha Rianna.,TimbangenSembiring.,Marhaposan Situmorang.,2018. Measurement and Analysis of Output Radiation Dose on X-Ray Device over 10 Years at Hospitals in Medan City.J. Aceh Phy. Soc. Vol. 7, No. 1 pp.1- 5, 2018

Laksono, M., Susila Wardaya, Wahyu Setia B., and Eko Hidayanto.Comparative Study Method of Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) for Determining Quality of Energy File Index 6 Mv And 10 Mv on Linac Aircraft. Periodical Physics ISSN: 1410 – 9662, Vol. 18, No. 1, January 2015, pp. 1-8

Sardari D., Maleki R., Samavat H.,hg Esmaeeli A., 2009, Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer code, Elselvier, Department of Medical Physics, Hamadan University of Medical Science, Hamadan, Iran. Reports of practical oncology and radiotherapy 15 (2010)64–68

Mariatul Khiftiyah, Eko Hidayanto dan Zaenal Arifin, 2014, Analysis Of Curves Percentage Depth Dose (PDD) And Profile Dose For The Field Of Simetri And Asymetric Radiation On Linear Accelerator (Linac) 6 And 10 Mv, Youngster Physics Journal, Vol. 3, No. 4, Oktober 2014, Hal 279- 286 ISSN : 2302 - 7371




DOI: https://doi.org/10.24815/jn.v18i2.11133

Refbacks